GateKeeper heated getter units
HGU p.1 describes point-of-use heated getter purification, including applications requiring CH4 and N2 impurity removal. Ar, He, Xe, N2 and H2 configurations are gas-specific. Nitrogen as a carrier gas is different from nitrogen as a target trace impurity; likewise, an EX hydrogen-removal statement cannot be transferred to a hydrogen carrier.
The p.2 PS2 and PS3 tables differ: PS2 lists several outlet specifications below 10 ppb and methane in nitrogen below 50 ppb, while PS3 has a separate table with below-1-ppb entries. Do not replace these conditions with one marketing-summary purity claim. Typical and maximum flows also differ.
On p.3, the PS4-MT3/MT15 life-status sensor statement applies to rare-gas and nitrogen configurations. It must not be extended to hydrogen or to every PS model.